Catalog No.
KAV13202
Description
PRINCIPLE OF THE ASSAY This assay employs the quantitative indirect enzyme immunoassay technique. Recombinant MPXV E8L antigen has been pre-coated onto a microplate. Standards or samples are pipetted into the wells and any Anti-MPXV E8L Human IgG present is bound by the immobilized protein. After washing away any unbound substances, a biotin-labeled Mouse Anti-Human IgG antibody is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-MPXV E8L Human IgG bound in the initial step. The color development is stopped and the intensity of the color is measured.
Applications
Used for the quantitative determination of Anti-MPXV E8L Human IgG concentration in serum and plasma.
Target
E8L
Detection method
Colorimetric
Sample type
Plasma, Serum
Assay type
Quantitative
Range
0.31 - 20 ng/mL
Sensitivity
0.30 ng/mL
Precision
Intra-Assay Precision (Precision within an assay): <10%
Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): <15%
Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
|
|
Intra-Assay Precision |
Inter-Assay Precision |
||||
|
Sample |
1 |
2 |
3 |
1 |
2 |
3 |
|
n |
16 |
16 |
16 |
24 |
24 |
24 |
|
Mean (ng/mL) |
9.9 |
2.4 |
1.2 |
9.7 |
2.4 |
1.1 |
|
Standard deviation |
0.3 |
0.1 |
0.1 |
0.3 |
0.1 |
0.1 |
|
CV (%) |
2.7 |
3.7 |
6.7 |
2.8 |
3.4 |
7.1 |
Recovery
80-120%
Shipping
2-8 ℃
Stability and Storage
When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%.
Note
For Research Use Only.