Elotuzumab ELISA Kit

Catalog #:   KDJ64001 Specific References (0) DATASHEET
Applications: Used for the quantitative determination of Elotuzumab concentration in serum and plasma.
Sample type: Plasma, Serum
Sensitivity: 151.74 ng/mL
Range: 156.25 - 10,000 ng/mL
Overview

Catalog No.

KDJ64001

Description

PRINCIPLE OF THE ASSAY This assay employs the quantitative competitive enzyme immunoassay technique. Recombinant Human SLAMF7 has been pre-coated onto a microplate. Standards or samples are premixed with biotin-labeled antibody and then pipetted into the wells. Elotuzumab in the sample competitively binds to the pre-coated protein with biotin-labeled Elotuzumab. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in inversely proportion to the amount of Elotuzumab bound in the initial step. The color development is stopped and the intensity of the color is measured.

Applications

Used for the quantitative determination of Elotuzumab concentration in serum and plasma.

Detection method

Colorimetric

Sample type

Plasma, Serum

Assay type

Quantitative

Range

156.25 - 10,000 ng/mL

Sensitivity

151.74 ng/mL

Precision

Intra-Assay Precision (Precision within an assay): <20%

Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.

Inter-Assay Precision (Precision between assays): <20%

Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.

 

Intra-Assay Precision

Inter-Assay Precision

Sample

1

2

3

1

2

3

n

16

16

16

24

24

24

Mean (ng/mL)

5736.0

1144.2

523.2

4825.0

1152.5

508.8

Standard deviation

586.4

155.0

82.1

407.4

108.4

93.9

CV (%)

10.2

13.5

15.7

8.4

9.4

18.5

 

Recovery

80-120%

Shipping

2-8 ℃

Stability and Storage

When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%.

Data Image
  • Experiment Example
    CALCULATION OF RESULTS Average the duplicate readings for each standard and sample. Construct a standard curve by plotting the mean absorbance for each standard on the Y-axis against the concentration on the X-axis and draw a best fit curve through the points on the graph. Do not include the blank in the standard curve. The data may be linearized by plotting the log of the Elotuzumab concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.
References
Datasheet

Document Download

Elotuzumab ELISA Kit.pdf

 

$ 1128
Product specifications
96 T 1128

Contact Information

Order: order@antibodysystem.com

Mail: support@antibodysystem.com

Distributor list

For research use only. Not for human or drug use.

Need help with your order?

Find out more about placing an order here

Elotuzumab ELISA Kit [KDJ64001]
Terms of sale Website terms of use Cookie policy Privacy
Copyright © 2025 AntibodySystem SAS. All Rights Reserved.            All Products are for Research Use Only