Catalog No.
KAB94001
Description
PRINCIPLE OF THE ASSAY This assay employs the quantitative sandwich enzyme immunoassay technique. Human Insulin has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Human Anti-Insulin Antibody will be captured by immobilized Human Insulin. After washing away any unbound substances, a biotin-labeled Human Insulin is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Human Anti-Insulin Antibody bound in the initial step. The color development is stopped and the intensity of the color is measured.
Applications
Used for the quantitative determination of Human Anti-Insulin Antibody concentration in serum and plasma.
Detection method
Colorimetric
Sample type
Plasma, Serum
Assay type
Quantitative (Sandwich)
Range
3.13 - 200 ng/mL
Sensitivity
2.13 ng/mL
Precision
Intra-Assay Precision (Precision within an assay): <10%
Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): <15%
Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Recovery
80-120%
Shipping
2-8 ℃
Stability and Storage
When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%.